4

ELECTRICAL CONNECTIONS

52

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

Control input active, modular I/Os

• U

int

= 24 VDC

• External contact open:

U

0,nom

= 22 V

External contact closed:
I

nom

= 4 mA

• Set switching point for detection "Contact open or closed:

Contact open (off): U

0

≤ 10 V with I

nom

= 1.9 mA

Contact closed (on): U

0

≥ 12 V with I

nom

= 1.9 mA

• X designates the connection terminals A or B, depending on the version of the signal

converter.

Control input passive, modular I/Os

• 3 V ≤ U

ext

≤ 32 VDC

• I

max

= 9.5 mA at U

ext

≤ 24 V

I

max

= 9.5 mA at U

ext

≤ 32 V

• Set switching point for detection "Contact open or closed:

Contact open (off): U

0

≤ 2.5 V with I

nom

= 1.9 mA

Contact closed (on): U

0

≥ 3 V with I

nom

= 1.9 mA

• X designates the connection terminals A or B, depending on the version of the signal

converter.

CAUTION!
Observe connection polarity.

Figure 4-26: Control input active C

a

1 Signal

Figure 4-27: Control input passive C

p

1 Signal

.book Page 52 Tuesday, August 18, 2009 4:51 PM