beautypg.com

Service – KROHNE OPTISONIC 6300 EN User Manual

Page 101

background image

SERVICE

7

101

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

• Remove the holding brackets 4 from the old electronics unit 5.
• Check compatibility between the removed and new electronics unit, by checking the power

voltage.

• Click the holding brackets 4 onto the new electronics unit and slide the new electronics unit

into the housing.

• Mount the small printed circuit board back onto the sensor driver board.
• Screw the electronics unit back to the housing.
• Click the display back into the holders.
• Close and lock the upper door, slide the metal slider upwards.
• Close and lock the lower door.
• Connect power.

Figure 7-8: Remove holding brackets

CAUTION!
First program the installation menu, refer to General instructions for programming on page 56
and check all important settings.

.book Page 101 Tuesday, August 18, 2009 4:51 PM