OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

Status output / limit switch

Status output / limit switch

Status output / limit switch

Status output / limit switch

Function and settings

Settable as automatic measuring range change, indicator for

direction of flow, overflow, error, operating point or empty pipe

detection

Valve control with activated dosing function

Status and/or control: ON or OFF

Operating data

Basic I/Os

Basic I/Os

Basic I/Os

Basic I/Os

Modular I/Os

Modular I/Os

Modular I/Os

Modular I/Os

Ex-i

Ex-i

Ex-i

Ex-i

Active

-

U

int

= 24 VDC

I ≤ 20 mA

open: I ≤ 0.05 mA

closed:

U

0, nom

= 24 V at

I = 20 mA

-

Passive

U

ext

≤ 32 VDC

I ≤ 100 mA

open:

I ≤ 0.05 mA at

U

ext

= 32 VDC

closed:

U

0, max

= 0.2 V at

I ≤ 10 mA

U

0, max

= 2 V at

I ≤ 100 mA

U

ext

= 32 VDC

I ≤ 100 mA

R

L, max

= 47 kΩ

open:

I ≤ 0.05 mA at

U

ext

= 32 VDC

closed:

U

0, max

= 0.2 V at

I ≤ 10 mA

U

0, max

= 2 V at

I ≤ 100 mA

-

NAMUR

-

Passive to

EN 60947-5-6

open: I

nom

= 0.6 mA

closed: I

nom

= 3.8 mA

Passive to

EN 60947-5-6

open: I

nom

= 0.43 mA

closed: I

nom

= 4.5 mA

U

I

= 30 V

I

I

= 100 mA

P

I

= 1 W

C

I

=10 nF

L

I

= 0 mH