ELECTRICAL CONNECTIONS

4

53

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

Control input active C

N

NAMUR, modular I/Os

• Connection in conformity with EN 60947-5-6
• Set switching point for detection "Contact open or closed:

Contact open (off): U

0, nom

= 6.3 V with I

nom

< 1.9 mA

Contact closed (on): U

0, nom

= 6.3 V with I

nom

> 1.9 mA

• Detection of cable break:

U

0

≥ 8.1 V with I ≤ 0.1 mA

• Detection of cable short circuit:

U

0

≤ 1.2 V with I ≥ 6.7 mA

• X designates the connection terminals A or B, depending on the version of the signal

converter.

CAUTION!
Observe connection polarity.

Figure 4-28: Control input active C

N

to NAMUR EN 60947-5-6

.book Page 53 Tuesday, August 18, 2009 4:51 PM