beautypg.com

GW Instek LCR-800 Series User Manual

Page 59

background image

INTERACE

117

/O_S_REJ

/O_S_OVER

The /O_S_REJ or /O_S_OVER signal
will go low when the secondary
measurement is over D_Max or
under D_Min, whilst in C/D , R/Q,
C/R or L/R mode. The signals will go
high at time T4

/O_EOM

The End of Measurement signal
becomes active low when the Bin
comparison/assignment has
completed. The signal goes high
after the next time I_E_TRIG is active
low.

Electrical
Characteristics

Output Characteristics

Output

Voltage

Signal Low

High

Max

current

/O_BIN1-BIN13

≤0.5V

+5V~+24V*

5mA*

/O_S_OVER

O_S_REJ

/O_P_OVER

/O_P_LO

/O_P_HI
Control Signals

/O_INDEX

/O_EOM

* Pull-up resistors R408~R427 must be replaced to
output greater than 5V.

LCR-800

User

Manual

118

Parameter Input

Signals

/I_E_TRIG

Measurement start signal. This
signal will trigger the LCR-800 to
start a measurement when the
signal is pulsed for at least 5us. It
is triggered by the falling edge of
the pulse.

/I_K_LOCK

The key lock signal disables the
panel keys when the signal is low,
and enables the panel keys when
the signal is high.

Electrical
Characteristics

Input Characteristics

Input Voltage

Input Current (Low)

Pull up voltage

Signal Low

High 5V 12V

/I_E_TRIG

≤1V

+5V~15V

5mA 12mA

/I_K_LOCK

≤1V

+5V~15V

5mA 12mA