Set bin menu – GW Instek LCR-800 Series User Manual
Page 31
BIN
FUNCTIONS
61
Set Bin Menu
Set Bin Menu Overview
Bin number
Displays the current bin.
Bin Settings
Configures the nominal value, nominal units and the
total amount of bins.
SORT BY
Chooses the primary or secondary
measurement to sort test results.
Tot_Bin
Configures the amount of sort bins.
R_Nom.Val
Q_Nom.Val
Sets the nominal value, depending on the
SORT BY R/Q settings.
Bin parameters
Configures the maximum and minimum sort limits for
the current bin.
Max: Ω+ Sets the maximum bin as an absolute
value.
Max: Ω- Sets the minimum bin as an absolute
value
Ω+: % Sets the maximum bin value as a positive
percentage offset from the nominal
value.
Ω- % Sets the minimum bin value as a
negative percentage offset from the
nominal value.
LCR-800
User
Manual
62
Q_Max
Sets the Maximum Q value for the
current bin.
Q_Min
Sets the minimum Q value for the
current bin.
Menu Keys
Scroll up through the menu items
Scroll down through the menu items
Clears the current bin settings.
Goes to the next bin.
Exit the menu
Bin Menu
Background
Before Bin Sorting, the measurement settings must
be configured
Panel operation
1. To access the handler menu, press MENU,
SORT, HANDLER.
2. The Handler menu appears.