BIN

FUNCTIONS

57

Bias Setting

Background

Set internal or external bias voltage.

Panel Operation

1. Use the arrow menu keys (F1/F2)

to move the cursor to INT.B or
EXT.B.

2. Press INT (F3) to use internal

biasing.

3. Press EXT (F4) to use external

biasing.

4. Use the arrow menu keys to

highlight OFF/ON.

5. Press ON (F3) to turn bias voltage

on.

6. Press OFF (F4) to turn bias voltage

off.

LCR-800

User

Manual

58

Range Setting

Background

The range can be selected from 1 to 4. Different
ranges should be used for different components
and component values and to ensure accurate
readings.

Component
Inductor

Capacitor

Resistor

Range1

1~16mH/f 1.6~25uF/f 6.25~100Ω

Range2

16~256mH/f 100~1600nF/f 0.1~1.6kΩ

Range3

256~4100mH/f 6.4~100nF/f 1.6~25.6kΩ

Range4

*

4.1~65H/f 400~6400pF/f 25.6~410kΩ

f = test frequency in kHz

*

This range is not used above 20 kHz

Panel Operation

1. Use the arrow menu keys (F1/F2)

to move the cursor to RANGE.

2. Press F3 (UP) to increase the range

or F4 (DOWN) to decrease the
range.

Constant Voltage Setting

Background

Constant voltage is usually used when a set
voltage is needed. For details about constant
voltage, see page 39.

Panel Operation

1. Use the arrow menu keys (F1/F2)

to move the cursor to C.V.