beautypg.com

GW Instek LCR-800 Series User Manual

Page 28

background image

BIN

FUNCTIONS

55

Select/Run Auto/Manu Sorting

Background

Set the test mode from manual to automatic.

Panel operation

1. Hold the START key for a few

seconds to toggle from automatic or
manual bin sorting.

The center of the screen will indicate if AUTO or
MANU mode is activated.

2. To test in MANU mode, press the

START key for each test. Testing
will begin automatically in AUTO
mode.

3. Results will be updated in the

display, depending on the settings.
Each time a test result has been
completed, an asterisk will appear
on the screen.

LCR-800

User

Manual

56

Voltage Setting

Background

Set the testing voltage.

Panel Operation

1. Use the arrow menu keys (F1/F2)

to move the cursor to V (Voltage)

2. Use the number pad to enter a voltage and

press ENTER to confirm.

1.000 V