BASIC

MEASUREMENT

35

Panel operation

1. Press the 7/Bias key on the number

pad to cycle from internal to
external bias. The bottom of the
screen will display internal or
external bias.

2. Press 8/ON/OFF to turn the bias

voltage on or off. The bottom of the
screen will display the internal or
external bias as on or off.

Set measurement frequency

Background

The measurement frequency, together with the
measurement voltage is used to define the
electrical characteristics of each measurement item.
Make sure the appropriate frequency is selected
according to the component characteristics.

The frequency range of each model is as follows:
100Hz~2kHz

LCR-816/826

12Hz~10kHz LCR-817/827
12Hz~100kHz LCR-819/829
12Hz~200kHz LCR-821

LCR-800

User

Manual

36

The LCR-821 can provide 504 different frequencies
with a 5 digit resolution including decimal places.
Any frequency can be keyed from the number pad,
and the closest available frequency (of 504) will be
selected automatically. The LCR-818/829 has 503
different frequencies and the LCR-817/827 and
LCR-816/826 have 489 and 245, respectively.

To calculate the different possible frequencies, use
the tables below.

LCR-821

Frequency range

Formula

n range

0.012 To

0.23077kHz

3kHz/n 13

to

250

0.23438 To 15kHz

60kHz/n

4 to 256

15.385

To 200kHz

200kHz/n

1 to 13

LCR-819/829

Frequency range

Formula

n range

0.012 To

0.23077kHz

3kHz/n 13

to

250

0.23438 To 15kHz

60kHz/n

4 to 256

15.385 To

100kHz 200kHz/n 2

to

13

LCR-817/827

Frequency range

Formula

n range

0.012 To

0.23077kHz

3kHz/n 13

to

250

0.23438 To 10kHz

60kHz/n

6 to 256

LCR-816/826

Frequency range

Formula

n range

0.10000 To 0.23077kHz 3kHz/n

13

to

30

0.23438 To 2kHz

60kHz/n

30 to 256

Panel operation

1. Press the -/FREQ key on the

number pad.

2. Enter the frequency using the numerical keys,

and then press ENTER.

1.0kHz