PROGRAMMING

93

Example MAIN:MODE:RQ<^END^M>

Sets the mode to R/Q (Resistance/Quality factor)

Query Syntax

MAIN:MODE?<^END^M>or<^J^M>

Return String

Current

measurement

mode

MAIN:MODE:RQ<^END>

R/Q

MAIN:MODE:CD<^END>

C/D

MAIN:MODE:LQ<^END>

L/Q

MAIN:MODE:LR<^END>

L/R

MAIN:MODE:ZQ<^END>

Z/Q

Query Example

MAIN:MODE?<^END^M>
MAIN:MODE:RQ<^END>

Returns the current measurement mode as R/Q

CIRCUIT

Command/Query

The mode command sets the equivalent circuit to series or parallel.

Syntax MAIN:CIRC:<^END^M>or<^J^M>

Parameter

Equivalent

Circuit

SERI

Series

PARA

Parallel

Example MAIN:CIRC:SERI<^END^M>

Sets the equivalent circuit to series

Query Syntax

MAIN:CIRC?<^END^M>or<^J^M>

Return String

Equivalent

circuit

MAIN:CIRC:SERI<^END> Series

LCR-800

User

Manual

94

MAIN:CIRC:PARA<^END> Parallel

Query Example

MAIN:CIRC?<^END^M>
MAIN:CIRC:PARA<^END>

Returns a parallel equivalent circuit as the current
setting.

FREQUENCY

Command/Query

Set or queries the test frequency.

Syntax MAIN:FREQ

<^END^M>or<^J^M>

Parameter

Frequency

(kHz)

0.01200~100.000 (7 characters, including a

decimal)

12 Hz~100kHz

Example MAIN:FREQ

0.01200<^END^M>

Sets the frequency to 12Hz (0.012 kHz)

Query Syntax

MAIN:FREQ?<^END^M>or<^J^M>

Return String

Frequency

MAIN:FREQ < variable ><^END>
(=0.01200~100.000)

Returns the test frequency
in kHz.

Query Example

MAIN:FREQ?<^END^M>
MAIN:FREQ 0.01200<^END>

Returns the current test frequency in kHz (12 Hz).

VOLTAGE

Command/Query

Set or queries the test signal voltage.

Syntax

MAIN:VOLT < variable ><^END^M>or<^J^M>

Parameter

< variable >

Test signal voltage