GW Instek LCR-800 Series User Manual
Page 51

PROGRAMMING
101
Parameter
< variable >
Nominal Value
-XXXXXXX ~
+XXXXXXX
Must be any 8 digit
character including a
decimal place and signage
(- or +).
+XXXXXX~-XXXXXX
(Mode dependant)
Example SORT:NOMV
–0.12345<^END^M>
Sets the nominal value to -0.12345
Query Syntax
SORT:NOMV?<^END^M>or<^J^M>
Return String
Value
SORT:NOMV < variable ><^END>
(
Returns the nominal value.
Query Example
SORT:NOMV?<^END^M>
SORT:NOMV 0.00200<^END>
Returns the nominal value 2Ω.
RECALL
Command/Query
Recall settings from one of 100 memory slots.
Syntax MEMO:RECA
Parameter
Memory
slot
1.00-100. (integer
values) 1-100
Note
Ensure the number has a total of 4 characters. If a
number does not use 4 characters, use a “.” and “0”
to “pad out” the number. Example 10 = 10.0
Example MEMO:RECA
100.
<^END^M>
Recalls saved settings from memory slot 100
Query Syntax
MEMO:NUMB?<^END^M>or<^J^M>
Return String
LCR-800
User
Manual
102
recall status
MEMO:NUMB
(
spsp
~100)
sp
=space character
OK. Returns the memory
slot used.
MEMO:RECA:EMPT<^END>
Not Ok. The memory slot
is empty, therefore no data
to recall.
Query Example
MEMO:NUMB?<^END^M>
MEMO:NUMB:100<^END>
Data was recalled from memory slot 100.
STORE
Command
Stores the current settings to one of 100 memory slots. A return
string will indicate the save slot used.
Syntax MEMO:STOR
Parameter
Memory
slot
1.00~100. (integer
values)
1-100
Note
Ensure the number has a total of 4 characters. If a
number does not use 4 characters, use a “.” and
additional zero’s (0) to “pad out” the number.
Example 10 = 10.0
Return String
Memory save slot
MEMO:STOR
(
spsp
~100)
sp
=space character
Returns the save slot used.
Example
MEMO:STOR 100.<^END^M>
MEMO:STOR 100<^END>
Data was saved to memory slot 100.