PROGRAMMING

109

SECONDARY FACTOR, PRIMARY UNIT,
SECONDARY UNIT

Measurement

Secondary measurement result is returned along with the primary
unit and secondary unit (C/R, L/R only). This measurement is the
second measurement displayed after measurements have been
started.

Return Syntax

MAIN:SECO <^END>

Test

result

Any 6 digit ASCII character including

sp

(+) or

– characters and a decimal point.

Secondary measurement
value

Primary

units

nF,Pf, uF

nanofarads, picofarads,
microfarads

Secondary

units

k,

sp

kΩ, Ω

Example

MAIN:SECO .0045nFk<^END>

The secondary measurement

result is .0045 with kΩ

as the unit. The primary unit is nF.

SECONDARY OVER, PRIMARY UNIT,
SECONDARY UNIT

Measurement

Secondary Over indicates that the secondary measurement exceeds
the measurement range of the LCR meter. Applicable for C/R &
L/R equivalent circuits with the display set to Value.

Return Syntax

SECO:OVER <^END>

Primary units

nF,Pf, uF, mH, Hsp (sp=space)

nanofarads, picofarads,
microfarads, millihenry,
henry

LCR-800

User

Manual

110

Secondary

units

k,

sp

kΩ, Ω

Example

SECO:OVER nFk<^END>

The secondary measurement result exceeds the range.
kΩ is the secondary unit and nF is the primary unit.

INITIATION HAS FINISHED (Initiate)

Command

Initiates the RS232 connection. A string is returned when the
initiation has been completed.

Syntax COMU:OVER<^END^M>or<^J^M>

Return String

Menu

level

COMU:OVER<^END>

Connection initiation
finished

Example COMU:OVER<^END^M>

COMU:OVER<^END>

Communication initiation has completed.

“RS232 ONLINE” will be displayed on the LCR-800
display panel.

OFF LINE

Command

Terminates the RS232 connection. A string is returned when the
initiation has been completed.

Syntax COMU:OFF.<^END^M>or<^J^M>

Return String

RS232

connection

COMU:OFF.<^END>

Terminated

Example COMU:OFF.<^END^M>

COMU:OFF.<^END>

The RS232 connection has been terminated.