Fixture Connection

Fixture structure

Background

The standard fixture is a four-wire type (Kelvin 4
wire). The outer terminals (Hforce and Lforce)
provide the current and the inner terminals
(Hsense and Lsense) measures the potential.

Diagram

Description HFORCE

Carries the signal current source.
Connected to the + side of the device
under test.

HSENSE

Together with Lsense, monitors the
Potential. Connected to the + side of
the device under test.

LSENSE

Together with Hsense, monitors the
Potential. Connected to the – side of
the device under test.

LFORCE

Accepts the signal current return.
Connected to the – side of the device
under test.

GND

If the test component has a large
metal area NOT connected to either
of the terminals, connect to the GND
input to minimize noise level.