GW Instek LCR-800 Series User Manual
Page 53

PROGRAMMING
105
COMU:MONO:819.<^END>
LCR-819
COMU:MONO:821.<^END>
LCR-821
Query Example
COMU:MONO<^END^M>
COMU:MONO:816.<^END>
The model number is LCR-816
ON-LINE
Query
The On-line function queries the RS232 connection status.
Query Syntax
COMU?<^END^M>or<^J^M>
Return String
connection
COMU:ON..<^END>
Connection on
COMU:OFF.<^END>
Connection off
Query Example
COMU?<^END^M>>
COMU:ON..<^END>
The RS232 connection is on.
MEASURE HOLD
Command
The Measure hold command is used to suspend measurement to
issue a new command when the LCR meter is busy. When the new
command is issued the Measure Recover command can be used to
resume measurement.
Syntax COMU:HOLD<^END^M>or<^J^M>
Example
COMU:HOLD<^END^M>
Measurement
is
suspended.
MEASURE RECOVER
Command
The Measure Recover command is used to resume measurements
after the Measure Hold command has been used.
Syntax COMU:RECO<^END^M>or<^J^M>
LCR-800
User
Manual
106
Example
COMU:RECO<^END^M>
Resume measurement. (recover measurement).
LEVEL DISPLAY
Command
Displays a menu level on the LCR-800 display. Returns the menu
level.
Syntax LEVE:
Parameter
Menu
Level
MAIN
Main
display
MENU
menu
display
PARA
Setting (Parameter) menu
SORT
Sort (Handler) menu
OFFS
Offset
menu.
Return String
level
LEVE:MAIN<^END>
Main display
LEVE:MENU<^END>
Menu display
LEVE:PARA<^END>
Setting (Parameter) menu
LEVE:SORT<^END>
Sort (Handler) menu
LEVE:OFFS<^END>
Offset menu.
Example
LEVE:MAIN<^END^M>
LEVE:MAIN<^END>
Set the display to the main display.
PRIMARY FACTOR
Measurement
Primary factor returns the primary measurement result, sans the
measurement unit. This measurement is the first measurement
displayed after measurements have been started.
Return Syntax
MAIN:PRIM