INTERACE

115

Pin11

/O_BIN_11

Go, Assigned BIN 11

Pin12

/O_BIN_12

Go, Assigned BIN 12

Pin13

/O_BIN_13

Go, Assigned BIN 13

Pin14

/O_S_OVER

No-Go/D or Q fail

Pin15

/O_P_OVER RLC

FAIL(O)

Pin16

GND GROUND

Pin17

VCC VCC

Pin18

GND GROUND

Pin19

/O_P_HI RLC

FAIL(O)

Pin20

/O_P_LO RLC

FAIL(O)

Pin21

/O_S_REJ

No-Go/D or Q fail

Pin22

/O_INDEX Data

acquisition

over,

OK to remove DUT(O)

Pin23

/O_EOM

End of Test(O)

Pin24

/I_E_TRIG Start

Measurement(I)

Pin25

/I_K_LOCK Panel

Lock


LCR-800

User

Manual

116

Signal Overview

Background

The signal overview section describes the
functions and the overall characteristics of the
signals used in the handler interface.

Parameter Output

Signals

/O_INDEX

The Index signal will become low
when the Analog measurement time
has completed. When the Index
signal is low, the test component can
be replaced with the next
component. The signal goes high
when the next trigger is active.

/O_BIN_1

~

/O_BIN_13

The Bin Go/No-Go signals go active
low when a successful comparison
has been made. For example if a
component is assigned to Bin_1,
/O_BIN_1 signal goes low until time
T4. All the remaining signals
(/O_BIN_2~/O_BIN_13) remain
high.

/O_P_HI

When the primary measurement is
higher than the MAX limit, O_P_HI
will go low until time T4.

/O_P_LO

When the primary measurement is
lower than the MIN limit, O_P_LO
will go low until time T4.

/O_P_OVER

When the primary measurement is
higher or lower than the MAX/MIN,
O_P_OVER will go low until time T4.