Signal overview – GW Instek LCR-800 Series User Manual
Page 58
INTERACE
115
Pin11
/O_BIN_11
Go, Assigned BIN 11
Pin12
/O_BIN_12
Go, Assigned BIN 12
Pin13
/O_BIN_13
Go, Assigned BIN 13
Pin14
/O_S_OVER
No-Go/D or Q fail
Pin15
/O_P_OVER RLC
FAIL(O)
Pin16
GND GROUND
Pin17
VCC VCC
Pin18
GND GROUND
Pin19
/O_P_HI RLC
FAIL(O)
Pin20
/O_P_LO RLC
FAIL(O)
Pin21
/O_S_REJ
No-Go/D or Q fail
Pin22
/O_INDEX Data
acquisition
over,
OK to remove DUT(O)
Pin23
/O_EOM
End of Test(O)
Pin24
/I_E_TRIG Start
Measurement(I)
Pin25
/I_K_LOCK Panel
Lock
LCR-800
User
Manual
116
Signal Overview
Background
The signal overview section describes the
functions and the overall characteristics of the
signals used in the handler interface.
Parameter Output
Signals
/O_INDEX
The Index signal will become low
when the Analog measurement time
has completed. When the Index
signal is low, the test component can
be replaced with the next
component. The signal goes high
when the next trigger is active.
/O_BIN_1
~
/O_BIN_13
The Bin Go/No-Go signals go active
low when a successful comparison
has been made. For example if a
component is assigned to Bin_1,
/O_BIN_1 signal goes low until time
T4. All the remaining signals
(/O_BIN_2~/O_BIN_13) remain
high.
/O_P_HI
When the primary measurement is
higher than the MAX limit, O_P_HI
will go low until time T4.
/O_P_LO
When the primary measurement is
lower than the MIN limit, O_P_LO
will go low until time T4.
/O_P_OVER
When the primary measurement is
higher or lower than the MAX/MIN,
O_P_OVER will go low until time T4.