PROGRAMMING

107

Test

result

Any 7 digit ASCII including

sp

(+) or –

characters and a decimal point.

Primary measurement
value

Example

MAIN:PRIM 32.705<^END>

The

primary

measurement is 32.705 (primary

measurement unit).

PRIMARY OV01

Measurement

Primary OV01 indicates that the primary measurement exceeds the
measurement range of the LCR meter. For example: If the
impedance of the DUT is less than the measurement range.

Return Syntax

PRIM:OV01<^END>

Example PRIM:OV01<^END>

Note, no units are returned

PRIMARY OVER SECONDARY OVER

Measurement

When both the primary and secondary factors exceed the range
(OVER), OVER will be returned.

Return Syntax

PRIM:OVER

Example PRIM:OVER

Note, no units are returned

SECONDARY FACTOR & PRIMARY UNIT

Measurement

Returns the secondary measurement results and the primary unit
(R/Q C/D L/Q only). This measurement is the second measurement
displayed after measurements have been started.

Return Syntax

MAIN:SECO <^END>

Test

Result

Any 6 digit ASCII character including

sp

(+) or

– characters and a decimal point.

Secondary measurement
value

LCR-800

User

Manual

108

Primary

unit

nF, pF, uF

nanofarads, picofarads,
microfarads

k

sp, spsp

(

sp

=

space character)

kΩ, Ω

mH, H

sp

millihenry,

henry

Example

MAIN:SECO .0045nF<^END>

The

secondary

measurement

is .0045 (D) and nF is

the primary measurement unit.

SECONDARY OVER & PRIMARY UNIT

Measurement

Secondary Over indicates that the secondary measurement exceeds
the measurement range of the LCR meter. The unit returned refers to
the primary measurement. Applicable for (R/Q, C/D, L/Q, Z/θ)
equivalent circuits.

Return Syntax

SECO:OVER<^END>

Primary

unit

nF, Pf, uF

nanofarads, picofarads,
microfarads

k

sp, spsp

(

sp

=

space character)

kΩ, Ω

mH, H

sp

millihenry,

henry

Example

SECO:OVER nF<^END>

The secondary measurement is OVER(exceeds range)
and nF is the primary measurement unit.