GW Instek LCR-800 Series User Manual
Page 54
PROGRAMMING
107
result
Any 7 digit ASCII including
sp
(+) or –
characters and a decimal point.
Primary measurement
value
Example
MAIN:PRIM 32.705<^END>
The
primary
measurement is 32.705 (primary
measurement unit).
PRIMARY OV01
Measurement
Primary OV01 indicates that the primary measurement exceeds the
measurement range of the LCR meter. For example: If the
impedance of the DUT is less than the measurement range.
Return Syntax
PRIM:OV01<^END>
Example PRIM:OV01<^END>
Note, no units are returned
PRIMARY OVER SECONDARY OVER
Measurement
When both the primary and secondary factors exceed the range
(OVER), OVER will be returned.
Return Syntax
PRIM:OVER
Example PRIM:OVER
Note, no units are returned
SECONDARY FACTOR & PRIMARY UNIT
Measurement
Returns the secondary measurement results and the primary unit
(R/Q C/D L/Q only). This measurement is the second measurement
displayed after measurements have been started.
Return Syntax
MAIN:SECO
Result
Any 6 digit ASCII character including
sp
(+) or
– characters and a decimal point.
Secondary measurement
value
LCR-800
User
Manual
108
unit
nF, pF, uF
nanofarads, picofarads,
microfarads
k
sp, spsp
(
sp
=
space character)
kΩ, Ω
mH, H
sp
millihenry,
henry
Example
MAIN:SECO .0045nF<^END>
The
secondary
measurement
is .0045 (D) and nF is
the primary measurement unit.
SECONDARY OVER & PRIMARY UNIT
Measurement
Secondary Over indicates that the secondary measurement exceeds
the measurement range of the LCR meter. The unit returned refers to
the primary measurement. Applicable for (R/Q, C/D, L/Q, Z/θ)
equivalent circuits.
Return Syntax
SECO:OVER
unit
nF, Pf, uF
nanofarads, picofarads,
microfarads
k
sp, spsp
(
sp
=
space character)
kΩ, Ω
mH, H
sp
millihenry,
henry
Example
SECO:OVER nF<^END>
The secondary measurement is OVER(exceeds range)
and nF is the primary measurement unit.