<6. PARAMETER DESCRIPTION>
6-44
IM 01E30D01-01EN
6.5.3 Alarms & Warning Messages
System Alarms (Device breakdown or inability to obtain correct measurements.)
Display unit/BRAIN
(60) content
Alarm countermeasure
message on display unit
NE-107
status
Alarm description
Countermeasure
10:μP Fault
Contact near office or
service center
F
Microprocessor (CPU) failure
Contact your nearest
Yokogawa office or service
center
11:EEPROM Fault
F
EEPROM failure
12:Sub uP Fault
F
Substitute mcroprocessor (CPU) failure
13:EX Pwr Fault
F
Extitation power circuit failure
14:A/D(S) Fault
F
A/D converter failure
15:A/D(I) Fault
16:Analog Fault
F
Analog circuit failure
17:Coil Open
F
Flowtube coil is open-circuit
18:Coil Short
F
Flowtube coil is short-circuit
19:Excite Error
F
Excitation circuit failure
20:Pulse Error
F
Discrepancy of pulse count
21:EEPROM Dflt
F
EEPROM default values
Process Alarms (Device is normal but process-related errors make correct measurement impossible.)
Display unit/BRAIN
(60) content
Alarm countermeasure
message on display unit
NE-107
status
Alarm description
Countermeasure
30:Sig Overflow
Check signal cableand
grounding
S
Input signal error
Carry out an investigation as follows:
• Check the signal cable for breakage.
• Check for stray currents in the fluid.
• Check the grounding.
• Check whether flowtube is filling with fluid.
• Check the electrode adhesion.
31:Empty Pipe
Fill flow tube with fluid
S
Flowtube is not filled with fluid Fill the flowtube with fluid.
32:H/L Alm
Check the flow rate and
setting value
–
Flow rate alarm for greater
than High limit value or less
than Low limit value.
Check the flow rate and setting value of
High limit and Low limit.
33:Adhesion Alm
Clean electrodes
S
Electrode adhesion alarm
Clean the electrodes.