beautypg.com

3 alarms & warning messages, Alarms & warning messages -44 – Yokogawa ADMAG AXR User Manual

Page 92

background image

<6. PARAMETER DESCRIPTION>

6-44

IM 01E30D01-01EN

6.5.3 Alarms & Warning Messages

System Alarms (Device breakdown or inability to obtain correct measurements.)

Display unit/BRAIN

(60) content

Alarm countermeasure

message on display unit

NE-107

status

Alarm description

Countermeasure

10:μP Fault

Contact near office or

service center

F

Microprocessor (CPU) failure

Contact your nearest

Yokogawa office or service

center

11:EEPROM Fault

F

EEPROM failure

12:Sub uP Fault

F

Substitute mcroprocessor (CPU) failure

13:EX Pwr Fault

F

Extitation power circuit failure

14:A/D(S) Fault

F

A/D converter failure

15:A/D(I) Fault
16:Analog Fault

F

Analog circuit failure

17:Coil Open

F

Flowtube coil is open-circuit

18:Coil Short

F

Flowtube coil is short-circuit

19:Excite Error

F

Excitation circuit failure

20:Pulse Error

F

Discrepancy of pulse count

21:EEPROM Dflt

F

EEPROM default values

Process Alarms (Device is normal but process-related errors make correct measurement impossible.)

Display unit/BRAIN

(60) content

Alarm countermeasure

message on display unit

NE-107

status

Alarm description

Countermeasure

30:Sig Overflow

Check signal cableand

grounding

S

Input signal error

Carry out an investigation as follows:

• Check the signal cable for breakage.

• Check for stray currents in the fluid.

• Check the grounding.

• Check whether flowtube is filling with fluid.

• Check the electrode adhesion.

31:Empty Pipe

Fill flow tube with fluid

S

Flowtube is not filled with fluid Fill the flowtube with fluid.

32:H/L Alm

Check the flow rate and

setting value

Flow rate alarm for greater

than High limit value or less

than Low limit value.

Check the flow rate and setting value of

High limit and Low limit.

33:Adhesion Alm

Clean electrodes

S

Electrode adhesion alarm

Clean the electrodes.