10.1.1 Zero Adjustment Using Display Unit Switches

This section describes the procedure for zero adjustment using the display unit switches. (For more details

regarding setting methods using these switches, refer to Chapter 5: Basic Operating Procedures.)

The parameters for zero adjustment are

B50/M10: Auto Zero Exe (and either of these can be used to carry

out this procedure). For more details regarding these parameters, refer to Chapter 6: Parameter Description.

The parameter

M10: Auto Zero Exe will be used in the following description.

Setting Mode

Major item

parameter

Setting mode

K:Diagnosis

M:Adjustment

N:Test

M10:Auto Zero Exe

No Execution

No Execution

Execution

M10:Auto Zero Exe

No Execution

Execution

No Execution

Major item

Sub-item

selection (D)

M10:Auto Zero Exe

Execution

Sub-item

Parameter

Search Mode

Parameter

Replacement

(execute) Mode

Start:

Major Item

Parameter

Search Mode

M:Adjustment

21:Adjustment 20mA

10:Auto Zero Exe

11:Autozero Time

SET

SET

INC

SET

SET

Once in Setting Mode, use the

INC

switch to

move the cursor to

M: Adjustment.

Touch the

SET

switch to access Sub-item

Parameter Search Mode.

Upon selection of

M: Adjustment, the cursor will

be positioned at

M10: Auto Zero Exe. (Sub-item

selection (D))

Touch the

SET

switch to access Parameter

Replacement Mode.

Touch the

INC

switch to move the cursor to

“Execution”.

Touch the

SET

switch to select the “Execution”.

In order to request confirmation, the entire display

flashes on and off. Touch the

SET

switch once

again at this time to fix selection of the automatic

zero adjustment function.

NOTE

When no operations are carried out for 20

seconds in the flashing state, the system will

automatically return to the Sub-item Parameter

Search Mode.

When the operations except

SET

are carried

out, the parameter cannot be set.