Yokogawa ADMAG AXR User Manual
Page 53

<6. PARAMETER DESCRIPTION>
6-5
IM 01E30D01-01EN
Item
Name
R/W
Data range
Units
Position
of decimal
point
Default value
(*): Indicated item
Description
Display unit
(BRAIN)
Display unit
/BRAIN
C43 Flow Decimal Pnt
(FLOW DECIMAL)
W
Auto
0
1
2
3
Auto (*)
Selects decimal point
position for the display unit’s
instantaneous flow rate. Linked
with B24.
C44 Velocity Check
(VELOCITY CHK)
R
0.000 to 99.999
m/s
3
Display of the span setting
using flow velocity (m/s).
C45 Density Unit
(DENSITY UNIT)
W
kg/m
3
lb/gal
lb/cf
kg/m
3
Sets units for density when
mass flow rate is selected.
C46 Mass Flow Density
(MASS DENSITY)
W
0 to 32000
C45
0 to 4
0
Sets density when mass flow
rate is selected (with units from
C45).
C50 User Span Select
(USER SPN SEL)
W
No
Yes
No
Selects whether or not special
units are used for flow rate
units.
C51 Flow User Unit
(FL USER UNIT)
W
8 alphanumeric
characters
Sets the special flow rate units.
C52 Flow User Span
(FL USER SPAN)
W
0.00001 to 32000
C51
0 to 4
100
Sets span when using special
flow rate units.
C60 —
(SELF CHECK)
R
Good
Error
See “6.5 Alarm Functions”.