[C41: Base Time Unit] Selection of time units for

the flow rate span

Refer to the description of parameter

B22.

[C42: Flow Span] Setting of the flow rate span

Refer to the description of parameter

B23.

[C43: Flow Decimal Pnt] Setting of the decimal

point position for the instantaneous flow rate

Refer to the description of parameter

B24.

[C44: Velocity Check] Display of the flow rate span

velocity

This parameter displays the flow rate span for the

maximum range in m/s units.

[C45: Density Unit] Setting of the density units for

mass flow rate

This parameter selects the units for density as

required when making settings using

C46: Mass

Flow Density.

[C46: Mass Flow Density] Setting of the density

for mass flow rate

This parameter is necessary in situations where

t, kg, g, klb or lb has been selected as the mass

unit in

B21/C40: Base Flow Unit. If a mass unit is

selected in

B21/C40: Base Flow Unit and a value

of 0 is set for this parameter, the setting alarm “

57:

Dens Set Err” will be displayed. In such a case,

ensure that the density is set correctly.

NOTE

Example : Since the water density is about

1000kg/m

3

, set the parameter below.

“kg/m

3

” is set for

C45: Density Unit,

“1000” is set for

C46: Mass Flow

Density.

Set the suitable value under the actual

use because the density changes by

temperature.

[C50: User Span Select] Selection of the use of

special flow rate units

This parameter selects whether or not special units

are used for instantaneous flow rate. Actual setting

of these units is carried out using

C51: Flow User

Unit and C52: Flow User Span.

[C51: Flow User Unit] Setting of the special flow

rate units

This parameter is used to select the special units

(up to maximum 8 characters in length). These

units are displayed when instantaneous flow rate

is selected in the Display Mode, and they are

displayed for

A20: FLOW RATE when BRAIN

communication is being carried out.

[C52: Flow User Span] Setting of the special flow

rate span

This parameter sets the special span to be

displayed for 100% output in the maximum range.

NOTE

Example : To set the special flow rate span to

100 dl/s

Since 100 dl (deci-liter)=10 l (liter),

“l (Liter)” is set for

B21/C40: Base

Flow Unit,

“/s” is set for

B22/C41: Base Time Unit,

“10” is set for

B23/C42: Flow Span,

“Yes” is set for

C50: User Span Select,

“dl/s” is set for

C51: Flow User Unit,

“100” is set for

C52: Flow User Span.

“100 dl/s” is indicated for 100% output

in the Display Mode.