A1.2.10 Lifetime Limits
The expected lifetime of the AXR is 10 years. The
reliability data listed the FMEDA report is only valid
for this period. The failure rates of the AXR may
increase sometime after this period. Reliability
calculations based on the data listed in the FMEDA
report for AXR lifetimes beyond 10 years may yield
results that are too optimistic, i.e. the calculated
Safety Integrity Level will not be achieved.
A1.2.11 Environmental Limits
The environmental limits of the AXR are specified in
the user’s manual IM 01E30D01-01E.
A1.2.12 Application Limits
The application limits of the AXR are specified in the
user’s manual IM 01E30D01-01E. If the flowmeter
is used outside of the application limits, the
reliability data listed in A1.2.9 becomes invalid.
A1.3 Definitions and
Abbreviations
A1.3.1 Definitions
Safety
Freedom from unacceptable
risk of harm
Functional Safety The ability of a system to carry
out the actions necessary
to achieve or to maintain a
defined safe state for the
equipment/machinery/plant/
apparatus under control of the
system
Basic Safety
The equipment must be
designed and manufactured
such that it protects against
risk of damage to persons
by electrical shock and other
hazards and against resulting
fire and explosion. The
protection must be effective
under all conditions of the
nominal operation and under
single fault condition
Verification
The demonstration for each
phase of the life-cycle that the
(output) deliverables of the
phase meet the objectives
and requirements specified by
the inputs to the phase. The
verification is usually executed
by analysis and/or testing
Validation
The demonstration that the
safety-related system(s) or
the combination of safety-
related system(s) and external
risk reduction facilities meet,
in all respects, the Safety
Requirements Specification.
The validation is usually
executed by testing
Safety Assessment The investigation to arrive at a
judgment -based on evidence-
of the safety achieved by
safety-related systems
Further definitions of terms used for safety
techniques and measures and the description of
safety related systems are given in IEC 61508-4.