beautypg.com

Important – Yokogawa ADMAG AXR User Manual

Page 148

background image

<11. MAINTENANCE>

11-8

IM 01E30D01-01EN

Sub-item

Parameter

Search Mode

Major item

Sub-item

selection (D)

F1110.EPS

K:Diagnosis

11:Adh Status

10:Adhesion Check

23:Empty Level

Now Adhesion Check

Executing

Fixed Current (4 mA)

0

100

Adhesion Diagnostic function is being executed

(about 5 minutes).

When Adhesion Diagnostic function has been

completed, the system automatically returns to the

sub-item selection screen (D).

IMPORTANT

In the event of a warning or alarm when

executing the adhesion diagnostic function, take

one of the following actions:

• Clean the electrodes, and then execute the

adhesion diagnostic function.

• Set the values of adhesion diagnostic level3

K15: Adhesion Level3 and level4 K16:

Adhesion Level4 bigger than the result of

the resistance value for adhesion diagnosis.

• Turn off the power to the flowmeter once,

and then turn on the power again.