<6. PARAMETER DESCRIPTION>
6-6
IM 01E30D01-01EN
(4) Item D (Menu D): Total Setting items
Menu D contains setting items such as the totalization scale and the forward/reverse totalized values.
Item
Name
R/W
Data range
Units
Position
of decimal
point
Default value
(*): Indicated item
Description
Display unit
(BRAIN)
Display unit
/BRAIN
D00 Total Set
(TOTAL SET)
D10 Total Unit
(TOTAL UNIT)
W
n Unit/P
u Unit/P
m Unit/P
Unit/P
k Unit/P
M Unit/P
Pulse/s
Pulse/s (*)
Selects the flow rate unit
per one pulse as used for
totalization display. Linked with
B30.
D11 Total Scale
(TOTAL SCALE)
W
0 to 32000
D10
(B30)
0 to 4
0 (*)
Sets the flow rate per one pulse
as used for totalization display.
Linked with B31.
D12 Total Decimal Pnt
(TL DECIMAL)
W
0
1
2
3
4
5
6
7
0
Selects position of decimal
point for totalization display
D13 Total Low Cut
(TOTAL LOWCUT)
W
0 to 100
%
0
3 %
Sets the range in vicinity of 0%
within which the totalization
display will be halted.
D14 Total Rate Check
(TL RATE CHK)
R
0 to 999999
Pulse/s
0 to 5
0.0
Displays the totalization rate in
“Pulse/s”
D20 Total Execution
(TOTAL EXEC)
W
Start
Stop
Preset Total
Preset Rev Total
Start
Executes “Start” or “Stop” of
the totalization function, or
executes “Preset Total” or
“Preset Rev Total”.
D21 Ttl Set Val Lower
(TL SET VAL L)
W
0 to 999999
0
0
Sets the totalization preset
value in the lower 6 digits of the
8-digit totalized value.
D22 Ttl Set Val Upper
(TL SET VAL U)
W
0 to 99
0
0
Sets the totalization preset
value in the upper 2 digits of
the 8-digit totalized value.
D23 Ttl Switch Lower
(TL SWITCH LO)
W
0 to 999999
0
0
Sets the totalization switch
value in the lower 6 digits of the
8-digit totalized value.
D24 Ttl Switch Upper
(TL SWITCH UP)
W
0 to 99
0
0
Sets the totalization switch
value in the upper 2 digits of
the 8-digits totalized value.
D30 Ttl User Select
(TL USER SEL)
W
No
Yes
No
Selects whether or not special
units are used as totalized
units.
D31 Ttl User Unit
(TL USER UNIT)
W
8 alphanumeric
characters
Sets the special totalized units.
D60 —
(SELF CHECK)
R
Good
Error
See “6.5 Alarm Functions”.