2 setting mode, 3 parameter setting procedure, Setting mode -5 – Yokogawa ADMAG AXR User Manual

Page 44: Parameter setting procedure -5 5.3.1

background image

<5. BASIC OPERATING PROCEDURES

(USING THE DISPLAY UNIT)

>

5-5

IM 01E30D01-01EN

5.2.2 Setting Mode

When the Setting Mode has been activated using the procedure from Section 5.2.1, parameters can be

selected for setting.

NOTE

If no operations are carried out for a period of 10 minutes in Setting Mode, the system will automatically

return to the Display Mode.

Format for Parameter Data
Depending on the type of parameter, data is formatted in one of the following three ways.

Format

Typical display

Content

(i) Selection-type

B21:Base Flow Unit

m

3

m

3

kl(Kiloliter)

B23:Flow Span

100 l/min

0000100. l/min

Rng:0.00001

32000

C10:Tag No

FI-1101

FI-1201

The desired data item is selected from a predefined list.

(ii) Numeric-type

B21:Base Flow Unit

m

3

m

3

kl(Kiloliter)

B23:Flow Span

100 l/min

0000100. l/min

Rng:0.00001

32000

C10:Tag No

FI-1101

FI-1201

Data is set using the values in each digit and using the decimal

point.

(iii) Alphanumeric-type

B21:Base Flow Unit

m

3

m

3

kl(Kiloliter)

B23:Flow Span

100 l/min

0000100. l/min

Rng:0.00001

32000

C10:Tag No

FI-1101

FI-1201

Data is composed using alphanumeric characters (in the case of tag

numbers, special units, and the like). With this format, setting can be

carried out using up to 16 of the characters shown below.

Regarding the alphanumeric-type format (iii), the following alphanumerics are displayed in the following sequence:

#%&*+-./0123456789:<>ABCDEFGHIJKLMNOPQRSTUVWXYZabcdefghijklmnopqrstuvwxyz[space]

5.3 Parameter Setting Procedure

Once the system is in Setting Mode, the parameters for setting can be selected. Parameters that are frequently

used have been grouped together in Easy Setup in Menu B. This section provides a description of the

parameter setting procedure using

B: Easy Setup and C: Basic Setup.

For more details regarding parameter content, please refer to Section 6: Parameter Description.

5.3.1 Setting Example for Selection-Type Data: Flow rate units

This example describes the setting of the flow rate units for the selection-type parameter

B21: Base Flow Unit

from m

3

to l (Liter).

Setting mode

P:Protect

B:Easy Setup

C:Basic Setup

SET

F0504-1.ai

Start:

Major Item

Parameter

Search Mode

Setting mode

Major item

parameter

Major Item Parameter Search Mode has been

accessed in this screen.

Touch the

SET

switch to access

B: Easy Setup.

NOTE

The ▲ and ▼ symbols to the left of the

parameters indicate that additional setting items

to those being currently displayed may also be

selected. Use the

INC

switch to cycle through

these items.