beautypg.com

Reference, Certifications, Emc compliance – Teledyne LeCroy WaveStation Waveform Generator Operators Manual User Manual

Page 80

background image

WaveStation

74

922700-00 Rev A

Reference

Certifications

This section certifies the instrument’s Electromagnetic Compatibility (EMC),
Safety and Environmental compliances.

EMC Compliance

EC

D

ECLARATION

O

F

C

ONFORMITY

-

EMC

The instrument meets intent of EC Directive 2004/108/EC for
Electromagnetic Compatibility. Compliance was demonstrated to the
following specifications as listed in the Official Journal of the European
Communities:

EN 61326-1:2006, EN 61326-2-1:2006 EMC requirements for electrical
equipment for measurement, control, and laboratory use.

1

Electromagnetic Emissions:

CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A

2 3

EN 61000-3-2:2006 Harmonic Current Emissions, Class A
EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers, Pst = 1

Electromagnetic Immunity:

EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV
vertical/horizontal coupling planes

4

EN 61000-4-3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80-1000
MHz; 3 V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz - 2.7 GHz

4

EN 61000-4-4:2004 Electrical Fast Transient/Burst, 1 kV on power supply
lines, 0.5 kV on I/O signal data and control lines

4

EN 61000-4-5:2006 Power line Surge, 1 kV AC Mains, L-N, L-PE, N-PE

4

EN 61000-4-6:2007 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz
- 80 MHz

4

EN 61000-4-11:2004 Mains Dips and Interruptions, 0%/1 cycle, 70%/25
cycles, 0%/250 cycles

4 5

1 To ensure compliance with all applicable EMC standards, high quality shielded interface
cables should be used.

2 Emissions which exceed the levels required by this standard may occur when the
instrument is connected to a test object.

3 This product is intended for use in nonresidential areas only. Use in residential areas may
cause electromagnetic interference.