Teledyne LeCroy WaveStation Waveform Generator Operators Manual User Manual
Page 48
WaveStation
42
922700-00 Rev A
Sweep Operation menu selections and parameters include SwpTime,
StopFreq, FreqSpan, StartFreq, MidFreq, Source, TrigOut, Linear/Log, and
Direction as follows:
SwpTime - This parameter sets the overall sweep time span in which the
frequency changes from specified start frequency to stop frequency
values.
StopFreq - Provide a specific frequency value where the sweep must
end.
FreqSpan - Provide an overall span or range in which the sweep must
remain.
StartFreq - Provide a specific frequency value where the sweep is to
begin.
MidFreq - Specify a frequency where the center of the sweep must
remain.
Source - Select Internal, External, or Manual. If External is selected, use
the Trig/Gate/Fsk/Burst connector on the Back Panel. You can also
select Manual and specify exact start and stop times.
TrigOut - Select Off or Open. Open sets trigger on the rising edge of the
waveform.
Linear/Log - Press this operation button to highlight and select either
Linear or Log spacing for your sweep waveform.
Direction - Press this operation button to toggle between an upward
(↑) or downward (↓) sweep direction for your waveform.