Operator’s Manual

922700-00 Rev A

55

B

Sync - Press this option button and an additional menu is provided
where you can turn the State of your synchronization On/Off and
use the Channel option button to send the synchronized signal to
CH1 or CH2.

All standard output functions (except DC and Noise) have a
corresponding Sync signal sent through the Sync Out connector on
the Back Panel. This can be disabled on certain settings as follows:

o Default - The Sync signal should be connected to the activated

Sync Out connector on the Back Panel. When the Sync Signal is
disabled, the output voltage of the [Sync] connector is set to low
level.

o Inverse - In this mode, the Waveform corresponding to the Sync

Signal does not inverse.

o Non-Modulated Waveforms - The Sync Signal reference is the

carrier signal on non-modulated waveforms.

o AM, FM, and PM - Internal modulating waveforms (AM, FM, and

PM) the Sync signal reference is the modulated signal

o ASK and FSK - The Sync Signal Reference is the keying Frequency.
o Sweep - Once the sweep starts, the Sync Signal becomes TTL Level

High. The Sync frequency equals the specific Sweep time.

o Burst - For the Burst, when the burst starts, the Sync Signal is

Level High.

o External Gated Burst - The Sync Signal follows the External Gated

Signal.

o Pulse - The Sync Signal is a Pulse Signal with a fixed positive pulse

width more than 50 ns.

ChCopy - Press this option button and an additional menu is
provided where you can port a particular channel's configuration to
another. Option buttons are labeled as CH1->CH2 and CH2->CH1
accordingly.

Make selections as desired and press the Done (to commit) or Cancel
option button and the main Util menu is shown.