78
L4400 User’s Guide
3
Operating and Programming
Figure 3-12. Channel Delay.
• You can set the channel delay to any value between 0 seconds and
60 seconds, with 1 ms resolution. You can select a different delay for
each channel.
• You can select a unique delay for every channel on the module.
• The channel delay is valid only while scanning. If no channels have
been assigned to the scan list, the specified channel delay is ignored
(no error is generated).
• The default channel delay is 0.0 seconds.
• A Factory Reset (
*RST
command) sets the channel delay to 0.0s.
The following command adds a 2- second channel delay to the specified
channels.
ROUT:CHAN:DELAY 2,(@1003,1013)
t
Scan List
t
Channel Delay
(0 to 60 seconds)
Ch 1
Ch 3
Ch 4
Ch 2
Ch 5
Ch 6
t
1
t
3
t
4
t
6
t
2
t
5