Calibration procedures, Adjustment is recommended, Agilent technologies calibration services – Agilent Technologies L4400 User Manual
Page 260: Calibration interval

248
L4400 User’s Guide
A
L4451A and L4452A Calibration Procedures
Calibration Procedures
This section contains performance verification and adjustment
(calibration) procedures for the L4400 Series L4451A 4- Channel Isolated
D/A Converter and L4452A Multifunction Module.
Closed-Case Electronic Calibration These instruments feature closed- case
electronic calibration. No internal mechanical adjustments are required.
The instruments calculate correction factors based upon the readings
from a separate DMM sent to the instruments during the calibration
procedure. The new correction factors are stored in non- volatile memory
until the next calibration adjustment is performed. Non- volatile EEPROM
calibration memory does not change when power has been off or after a
remote interface reset.
Agilent Technologies Calibration Services
When your instruments are due for calibration, contact your local Agilent
Service Center for low- cost recalibration. The L4451A and L4452A are
supported on automated calibration systems which allow Agilent to
provide this service at competitive prices.
Calibration Interval
The instruments should be calibrated on a regular interval determined by
the accuracy requirements of your application.
A 1- year interval is adequate for most applications. Accuracy
specifications are warranted only if adjustment is made at regular
calibration intervals. Accuracy specifications are not warranted beyond
the 1- year calibration interval. Agilent does not recommend extending
calibration intervals beyond 2 years for any application.
Adjustment is Recommended
Specifications are only guaranteed within the period stated from the last
adjustment. Whatever calibration interval you select, Agilent recommends
that complete re- adjustment should always be performed at the
calibration interval. This will assure that the L4451A and L4452A will
remain within specification for the next calibration interval. This criteria
for re- adjustment provides the best long- term stability.
Performance data measured during Performance Verification Tests does
not guarantee the instruments will remain within these limits unless the
adjustments are performed.
Use the Calibration Count feature (see
) to verify that all
adjustments have been performed.