Y1152a switch control – Agilent Technologies L4400 User Manual
Page 179
L4445A Microwave Switch/Attenuator Driver
7
L4400 User’s Guide
167
16 Conductor Cable
9 Conductor Cable
Y1152A Switch Control
All channels are driven in
PAIRed
mode.
* For switches connected to SW1, note the path closed is accomplished with the
ROUTe:OPEN
command.
Item
Description
Example Part Numbers
Cable Type
16 conductor ribbon cable, 0.050" pitch, 26 or 28
AWG stranded
3M 3801/16 (26 AWG)
3M 3365/16 (28 AWG)
Y1152A
Connector
10 pin socket connector, 0.1" x 0.1" pin grid, IDC
termination, center polarizing key
3M P/N 89116-0101
AMP P/N 76288-3
Switch
Connector
16 pin socket connector, 0.1" x 0.1" pin grid, IDC
termination, center polarizing key
3M P/N 89116-0101
AMP P/N 76288-3
Cable
Wiring
Y1152A connector pin 1 to switch connector pin 1
Item
Description
Example Part Numbers
Cable Type
9 conductor ribbon cable, 0.050" pitch, 26 or
28 AWG stranded
3M 3801/09 (26 AWG)
3M 3365/09 (28 AWG)
Y1150A Connector
10 pin socket connector, 0.1" x 0.1" pin grid,
IDC termination, center polarizing key
3M P/N 89110-0101
AMP P/N 76288-1
Switch Connector
9 pin D-sub male, IDC termination, without
threaded insert
3M P/N 8209-6000
AMP P/N 747306-4
Cable Wiring
Y1152A socket connector pin 1 to switch
D-sub connector pin 1
(Note: pin 10 of Y1152A connector not
used)
Path closed*
Path open*
SW1 Path1
ROUT:OPEN (@xx01)
ROUT:CLOS (@xx01)
SW1 Path2
ROUT:OPEN (@xx02)
ROUT:CLOS (@xx02)
SW1 Path3
ROUT:OPEN (@xx03)
ROUT:CLOS (@xx03)
SW1 Path4
ROUT:OPEN (@xx04)
ROUT:CLOS (@xx04)
SW1 Path5
ROUT:OPEN (@xx05)
ROUT:CLOS (@xx05)
SW1 Path6
ROUT:OPEN (@xx06)
ROUT:CLOS (@xx06)
State A
State B
SW2
ROUT:OPEN (@xx07)
ROUT:CLOS (@xx07)
SW3
ROUT:OPEN (@xx08)
ROUT:CLOS (@xx08)