L4400 User’s Guide
3
Operating and Programming
instrument must be used when connecting the L4400 instrument directly
(without a switch or router) to the computer. See “Connecting the LAN
Cables” in Chapter 2 for more information.
L4400 Series Channel Addressing Scheme
The channel addressing scheme for the L4400 series LXI instuments uses
the form 1ccc where ccc is the three- digit channel number. Following are
examples of the scheme. Refer to the individual instrument chapters for
more information on channel numbering.
Table 3-2. L4400 Series Channel Addressing Examples.
Channel Number
Definition
1014
Channel 14 on Bank 1 of the L4421A multiplexer module.
1921
Channel 921 (analog bus relay for ABUS 1) on Bank 2 of the
L4421A multiplexer module or on Matrix 2 of the L4433A matrix
module.
1304
Row 3 / column 4 on Matrix 1 of the L4433A matrix module
(2-wire mode).
1437
Row 3 / column 7 on Matrix 4 of the L4433A matrix module
(1-wire mode).
1201
Channel 201 on Bank 2 of the L4450A digital I/O module.
Introduction to the SCPI Command Language
The functions of the L4400 series instruments are programmed using
SCPI (Standard Commands for Programmable Instruments) commands.
The L4400 Programmer’s Reference located on the L4400 LXI Class C
Instruments Product Reference CD- ROM (p/n 34989- 13601), contains a
complete description of each instrument’s command set.
SCPI is an ASCII- based instrument command language designed for test
and measurement instruments. SCPI commands use a hierarchical
structure, also known as a tree system. In this system, associated
commands are grouped together under a common node or root, thus
forming subsystems. A portion of the ROUTe subsystem is shown below
to illustrate the tree system.
ROUTe
:MONitor
[:CHANnel]:ENABle
[:CHANnel]:ENABle? (@