Page 261

background image

L4451A and L4452A Calibration Procedures

A

L4400 User’s Guide

249

Time Required for Calibration

The L4451A and L4452A can be automatically calibrated under computer
control. With computer control you can perform the complete calibration
procedures and performance verification tests in less than 30 minutes
once the instruments are warmed- up (see “L4451A and L4452A
Performance Test Considerations”).

Automating Calibration Procedures

You can automate the complete verification and adjustment procedures
outlined in this chapter. You can program the instrument configurations
specified for each test and then enter readback verification data into a
test program and compare the results to the appropriate test limit
values.

The instruments must be unsecured prior to initiating the calibration
procedures (see “Calibration Security”).

Recommended Test Equipment

The test equipment recommended for the performance verification and
adjustment procedures is listed in Table A- 1. If the exact instrument is
not available, substitute calibration standards of equivalent accuracy.

Table A-1. Recommended Test Equipment

Application

Recommended Equipment

Accuracy Requirements

Analog Output
L4451A

Agilent 34401A, 34410A, or
34411A

<1/5 L4451A 24-hour voltage /
current specification

Analog Output
L4452A

Agilent 34401A, 34410A, or
34411A

<1/5 L4452A 24-hour voltage /
current specification