L4451A and L4452A Calibration Procedures

A

L4400 User’s Guide

263

4

Connect channel 6 of the L4452A to the DMM as shown in Figure A- 3.
Set the DMM to measure DC volts.

5

Set the first DAC calibration point using the command:


CALibration:BEGin:VOLTage 1, (@)


For channel 6, the command is executed as:


CALibration:BEGin:VOLTage 1, (@1006)

6

Measure the DAC output on the DMM. Subtract the offset voltage
measured in Step 2 from the reading. Write the result to the DAC
using the command:


CALibration:POINt?


Note the DAC output which is now calibration point 2. Subtract the
offset voltage from the new DMM reading and write the value to the
DAC using:


CALibration:POINt?

7

Connect channel 7 of the L4452A to the DMM as shown in Figure A- 3.
Repeat Steps 5 and 6 for channel 7.