L4451A and L4452A Calibration Procedures
A
L4400 User’s Guide
263
4
Connect channel 6 of the L4452A to the DMM as shown in Figure A- 3.
Set the DMM to measure DC volts.
5
Set the first DAC calibration point using the command:
CALibration:BEGin:VOLTage 1, (@
For channel 6, the command is executed as:
CALibration:BEGin:VOLTage 1, (@1006)
6
Measure the DAC output on the DMM. Subtract the offset voltage
measured in Step 2 from the reading. Write the result to the DAC
using the command:
CALibration:POINt?
Note the DAC output which is now calibration point 2. Subtract the
offset voltage from the new DMM reading and write the value to the
DAC using:
CALibration:POINt?
7
Connect channel 7 of the L4452A to the DMM as shown in Figure A- 3.
Repeat Steps 5 and 6 for channel 7.