L4450A 64-Bit Digital I/O with Memory and Counter

8

L4400 User’s Guide

213

Start the capture. The

SENSe:DIGital:MEMory:STARt

command sets the

channel to begin the data capture. The capture begins when the handshake
occurs.

Check the status of the transfer. You can use the

SENSe:DIGital:MEMory:DATA:POINts?

query to return the number of

samples currently in memory.

Retrieve the captured data. Set the desired memory retrieval format using
the

SENSe:DIGital:DATA:FORMat

command. You can set the memory to be

read as either

LIST

or

BLOCk

. The

LIST

parameter (default) returns the

data as comma separated ASCII values.

BLOCk

returns the data in

IEEE- 488 block format.

Before you can read the data in memory, you must stop the memory
operations using the

SENSe:DIGital:MEMory:ENABle OFF

command.

Read all the captured data using the

SENSe:DIGital:MEMory:DATA:ALL?

query. This performs a non- destructive read of all data in the bank’s
memory.

To read specific captures, use the

SENSe:DIGital:MEMory:DATA?

form of

the command. This command takes index and count parameters to specify
which data to retrieve. The oldest data in memory has an index of ‘0’. The
count specifies the number of samples to read. count + index must be less
than the number of captured points.

Both these data reads are non- destructive to the bank memory. To clear
the memory for new data, send the

SENSe:DIGital:MEMory:CLEar

command.

Interrupt Lines

Each bank has an interrupt line that can be used with memory input or
output operations. When a bank is set to input data, the interrupt line is
an output. When a bank is set to output data, the interrupt line is set to
be an input. You can set the polarity of the interrupt line for input and
output operations using the

CONFigure:DIGital:INTerrupt:POLarity

command.

You can configure the interrupt line drive mode, output drive level, and
input threshold. These parameters are set for both the handshake lines
and interrupt line on a bank. See

page 204

for details about setting these

parameters.

Memory Output Operations

For memory output operations, the interrupt line is sensed and can be
used to start or stop memory output operations. This provides a hardware
means to control the data output.