The following program segment selects the alarm source and configures
the instrument (L4450A) to execute the sequence named “TOTAL_1”
when an alarm (count of 50) is reported on Alarm 1. The Monitor mode
is used to evaluate alarm conditions on the selected channel.

ROUT:SEQ:DEF TOTAL_1,"SYST:DEL 1000"

CALC:LIM:UPP 10.25,(@1301)

CALC:LIM:UPP:STAT ON,(@1301)

OUTP:ALARM1:SOUR (@1301)

ROUT:MON:CHAN (@1301)

ROUT:MON:CHAN:ENAB ON, (@1301)

ROUT:SEQ:TRIG:SOUR TOTAL_1,ALAR1

ROUT:MON:STAT ON

INIT

Deleting Sequences

You can delete sequences to free up space in memory previously
allocated for the sequence.

If you attempt to delete a sequence while it is executing, an error

will be generated. To abort a sequence execution, use the

ROUTe:SEQuence:ABORt

command or a

Device Clear

.

Deleting a sequence will remove its association with an alarm if used

(see

“Executing a Sequence on an Alarm Condition"

on page 94 for

more information).

The following command deletes the sequence named “TOTAL_1”.

ROUT:SEQ:DEL TOTAL_1

The following command deletes all sequences from memory.

ROUT:SEQ:DEL:ALL

Reading the List of Stored Sequences

You can read the names of all sequences currently stored in memory.

When stored in memory, the user- defined sequence names are

converted to all uppercase letters. For example when stored, “Total_1”
is converted to “TOTAL_1”.

Up to 500 unique sequences can be stored in non- volatile memory.

Each sequence is limited to 1024 bytes.

The following command returns a comma- separated list of sequence
names currently stored.

ROUT:SEQ:CAT?