beautypg.com

Simple switch control – Agilent Technologies L4400 User Manual

Page 156

background image

144

L4400 User’s Guide

7

L4445A Microwave Switch/Attenuator Driver

Simple Switch Control

All examples in this chapter make reference to SCPI commands for switch
control. The L4445A commands are summarized in Table 7- 1. For more
information, refer to the Programmer’s Reference help file included on the
Product Reference CD- ROM (p/n 34989- 13601).

The switches and attenuators are designed to respond to the SCPI

ROUTe:CLOSe

and

ROUTe:OPEN

commands. For example, to open and close

a switch attached to channel 1 on bank 1 of the second remote module
attached to the L4445A, you could use the following commands (rem = 2,
channel = 01).

ROUT:OPEN (@1201)

ROUT:CLOS (@1201)

Before you can close or open a switch, however, several other parameters
must be configured. Each distribution board has a set of factory default
parameters designed to support the type of switches intended to be
present. These defaults are described in more detail on

page 152

.

Additionally, the drive current source must be selected and configured.

The following commands show a simple sequence controlling channel 1 of
an Agilent N1810 switch (installed on a Y1150A distribution board) of the
third remote module attached to a L4445A (rem = 3, channel = 01). Note
that ALL 34945EXT modules are reset by the first command shown.

SYST:RMOD:RES 1

ROUT:RMOD:BANK:PRESET BANK1,(@1300)

ROUT:CHAN:DRIV:CLOS:DEF (@1301)

ROUT:RMOD:DRIV:SOUR INT,(@1300)

ROUT:OPEN (@1301)

<-- other commands -->

ROUT:CLOS (@1301)

In the example above, the

SYSTem:RMODule:RESet

command resets the

module and disables all drive currents. The next command loads the
factory default settings for the distribution board (Y1150A) used to
support the Agilent N1810 switch. The default state of switch closed is
then configured. When the drive source is set to internal (third remote
module only), the switch assumes its default closed state. The configured
switch may now be controlled using the

ROUTe:OPEN

and

ROUTe:CLOSe

commands.

These commands and settings are described in more detail later in this
chapter and in the Programmer’s Reference Help file.

N O T E

You must turn off the channel drive before sending the

ROUTe:RMODule:BANK:PRESet

command. Once configured, turn the

channel drive back on
(

ROUTe:RMODule:DRIVe:SOURce:IMMediate

).