FLUID COMPONENTS INTERNATIONAL LLC http://www.fluidcomponents.com

Model GF03 Flow Meter

2 8

Doc. No. 06EN003352 Rev. B

Application fo the correction equation will be illustrated in the following example. Through
analysis of the process flow stream it is determined that the flow rate must be multiplied by
1.056 to output the desired value.

The correction equation takes the following form:

m

new

= 0 + (1.056 x m

o

) + (0 x m

o

2

)+ (0 x m

o

3

)

Therefore m

new

= 1.056 x m

o

Where

F

1

= 0

F

2

= 1.056

F

3

= 0

F

4

= 0

A more complicated situation would be where the multiplication factor varies with flow rate.
Figure 21 shows the variation of desired flow rate versus the measured flow rate.

Is the

correction

equation

enabled?

Apply correction

equation

Corrected flow rate

Apply other

corrections (see

Aux. Input)

Yes

C00247-2

Manipulate

analog outputs

and relays

Measured flow

rate

NO

Measured Flow Rate

Desired

Flow Rate

0

10

20

30

40

50

0

20

40

60

C00248-1

Figure 21 - Desired Flow Rate versus Measured Flow Rate

Figure 20 - Correction Factor