beautypg.com

Fluid Components International 8-66B_12-64B Series Manual Installation User Manual

Installation

background image

Doc. No. 06EN003264 Rev. A

2 - 1

Models 12-64/8-66 Basic Switch

CHAPTER 2 - INSTALLATION

FLUID COMPONENTS INTL

2. Installation

Receiving/Inspection

Unpack carefully.

Verify that all items in the packing list are received and are correct.

Inspect all instruments for damage or contaminants prior to installation.

If the above three items are satisfactory, proceed with the installation. If not, then stop and contact a customer
service representative.

Packing/Shipping/Returns

These issues are addressed in Appendix C - Customer Service

Required Materials

Appropriate wire, cable, conduit, and a mating interface on the vessel.

Note:

Potting Y's for all the interconnecting wires are recommended when installing the instrument. Other
requirements may vary based on local wiring codes. An appropriate mating surface on the vessel is
also required.

Pre-Installation Procedure

Warning:

Only qualified personnel should install this instrument. Install and follow safety procedures in
accordance with the current National Electrical Code. Ensure that power is off during
installation. Any instances where power is applied to the instrument will be noted in this manual.
Where the instructions call for the use of electrical current, the operator assumes all responsibility
for conformance to safety standards and practices.

Caution:

The control circuit contains electrostatic discharge (ESD) sensitive devices. Use standard ESD
precautions when handling the control circuit. See below for ESD details.

The instrument is not designed for weld-in-place applications. Never weld to process connection
or a structural support.

Damage resulting from moisture penetration of the local enclosure is not covered by product
warranty.

Use Standard ESD Precautions

Use standard ESD precautions when opening an instrument enclosure or handling the level transmitter. FCI
recommends the use of the following precautions: Use a wrist band or heel strap with a 1 megohm resistor connected
to ground. If the instrument is in a shop setting there should be static conductive mats on the work table and floor
with a 1 megohm resistor connected to ground. Connect the instrument to ground. Apply antistatic agents to hand
tools to be used on the instrument. Keep high static producing items away from the instrument such as non-ESD
approved plastic, tape and packing foam.