Error Codes
Code Number
Description
Correction
Warnings
0001
Serial number not present
Hardware serial number has become inoperative –
system performance will not be influenced.
0010
Signal Strength is below Signal Strength Cutoff
entry
Low signal strength is typically caused by one of the
following:
» Empty pipe
» Improper programming/incorrect values
» Improper transducer spacing
» Non-homogeneous pipe wall
0011
Measured Speed of Sound in the liquid is
greater than ±10% different than the value
entered during meter setup
Verify that the correct liquid was selected in the
BASIC menu.
Verify that pipe size parameters are correct.
Class C Errors
1001
System tables have changed
Initiate a meter RESET by cycling power or by selecting
SYSTEM RESET in the SEC MENU.
1002
System configuration has changed
Initiate a meter RESET by cycling power or by selecting
SYSTEM RESET in the SEC MENU.
Class B Errors
3001
Invalid hardware configuration
Upload corrected file.
3002
Invalid system configuration
Upload corrected file.
3003
Invalid strategy file
Upload corrected file.
3004
Invalid calibration data
Re-calibrate the system.
3005
Invalid speed of sound calibration data
Upload new data.
3006
Bad system tables
Upload new table data.
Class A Errors
4001
Flash memory full
Return unit to factory for evaluation
Revised 5-25-2009
Table 21: Flow meter error codes