Error Codes

Code Number

Description

Correction

Warnings

0001

Serial number not present

Hardware serial number has become inoperative –

system performance will not be influenced.

0010

Signal Strength is below Signal Strength Cutoff

entry

Low signal strength is typically caused by one of the

following:

» Empty pipe

» Improper programming/incorrect values

» Improper transducer spacing

» Non-homogeneous pipe wall

0011

Measured Speed of Sound in the liquid is

greater than ±10% different than the value

entered during meter setup

Verify that the correct liquid was selected in the

BASIC menu.
Verify that pipe size parameters are correct.

Class C Errors

1001

System tables have changed

Initiate a meter RESET by cycling power or by selecting

SYSTEM RESET in the SEC MENU.

1002

System configuration has changed

Initiate a meter RESET by cycling power or by selecting

SYSTEM RESET in the SEC MENU.

Class B Errors
3001

Invalid hardware configuration

Upload corrected file.

3002

Invalid system configuration

Upload corrected file.

3003

Invalid strategy file

Upload corrected file.

3004

Invalid calibration data

Re-calibrate the system.

3005

Invalid speed of sound calibration data

Upload new data.

3006

Bad system tables

Upload new table data.

Class A Errors
4001

Flash memory full

Return unit to factory for evaluation

Revised 5-25-2009

Table 21: Flow meter error codes