Min Flow is the minimum volumetric flow rate setting entered to establish filtering parameters. Volumetric entries will be in

the flow rate units. For unidirectional measurements, set Min Flow to zero. For bidirectional measurements, set Min Flow to the

highest negative (reverse) flow rate expected in the piping system.
Max Flow is the maximum volumetric flow rate setting entered to establish filtering parameters. Volumetric entries will be in

the flow rate units. For unidirectional measurements, set Max Flow to the highest (positive) flow rate expected in the piping

system. For bidirectional measurements, set Max Flow to the highest (positive) flow rate expected in the piping system.
Low Flow Cutoff is provided to allow very low flow rates (that can be present when pumps are off and valves are closed) to be

displayed as zero flow. Typical values that should be entered are between 1.0…5.0% of the flow range between

Min Flow and Max Flow.
Low Signal Cutoff is used to drive the flow meter and its outputs to the value specified in the Substitute Flow field when

conditions occur that cause low signal strength. A signal strength indication below 5 is generally inadequate for measuring

flow reliably, so generally the minimum setting for low signal cutoff is 5. A good practice is to set the low signal cutoff at

approximately 60…70% of actual measured maximum signal strength.

OTE:

N

The factory default low signal cutoffis five.

If the measured signal strength is lower than the low signal cutoff setting, a Signal Strength too Low highlighted in red will

become visible in the text area to the left in the Data Display screen until the measured signal strength becomes greater than

the cutoff value.
Signal strength indication below two is considered to be no signal at all. Verify that the pipe is full of liquid, the pipe size and

liquid parameters are entered correctly, and that the transducers have been mounted accurately. Highly aerated liquids will

also cause low signal strength conditions.
Substitute Flow is a value that the analog outputs and the flow rate display will indicate when an error condition in the

flow meter occurs. The typical setting for this entry is a value that will make the instrument display zero flow during an

error condition.
Substitute flow is set as a percentage between Min Flow and Max Flow. In a unidirectional system, this value is typically set

to zero to indicate zero flow while in an error condition. In a bidirectional system, the percentage can be set such that zero

is displayed in an error condition. To calculate where to set the Substitute Flow value in a bidirectional system, perform the

following operation:

100

100

-

-

=

Maximum Flow

Substitute Flow

Maximum Flow

Minimum Flow

Entry of data in the Basic and Flow tabs is all that is required to provide flow measurement functions to the flow meter. If the

user is not going to utilize input/output functions, press Download to transfer the configuration to the flow meter instrument.

When the configuration has been completely downloaded, turn the power to the meter off and then on again to guarantee

the changes take effect.