Allows adjustments to be made to the speed of sound value, shear or transverse wave, for the pipe wall. If the UNITS value was

set to ENGLSH, the entry is in fps (feet per second). METRIC entries are made in mps (meters per second).
If a pipe material was chosen from the PIPE MAT list, a nominal value for speed of sound in that material will be automatically

loaded. If the actual sound speed is known for the application piping system and that value varies from the automatically

loaded value, the value can be revised.
If OTHER was chosen as PIPE MAT, then a PIPE SS must also be entered.
Pipe Roughness

PIPE R — Pipe Material Relative Roughness (Value)

Unitless Value

The flow meter provides flow profile compensation in its flow measurement calculation. The ratio of average surface

imperfection as it relates to the pipe internal diameter is used in this compensation algorithm and is found by using the

following formula:

 

 

   

 

 

 

 

   

 

Linear RMS Measurement of the Pipes Internal Wall Surface

Inside Diamet

Pipe

er of the Pip

R

e

=

If a pipe material was chosen from the PIPE MAT list, a nominal value for relative roughness in that material will be

automatically loaded. If the actual roughness is known for the application piping system and that value varies from the

automatically loaded value, the value can be revised.
Liner Thickness

LINER T — Pipe Liner Thickness (Value)

ENGLSH (Inches)

METRIC (Millimeters)

If the pipe has a liner, enter the pipe liner thickness. Enter this value in inches if ENGLSH was selected as UNITS; in millimeters if

METRIC was selected.
Liner Material

LINER MA — Pipe Liner Material (Choice)

Liner Type - (If a LINER Thickness was selected)

Tar Epoxy

TAR EPXY

HD Polyethylene

HDPE

Rubber

RUBBER

LD Polyethylene

LDPE

Mortar

MORTAR

Teflon (PFA)

TEFLON

Polypropylene

POLYPRO

Ebonite

EBONITE

Polystyrene

POLYSTY

Other

OTHER

This list is provided as an example. Additional materials are added periodically. Select the appropriate material from the list or

select OTHER if the liner material is not listed.
Liner Sound Speed

LINER SS — Speed of Sound in the Liner (Value)

ENGLSH (Feet per Second)

METRIC (Meters per Second)

Allows adjustments to be made to the speed of sound value, shear or transverse wave, for the pipe wall. If the UNITS value was

set to ENGLSH, the entry is in fps (feet per second). METRIC entries are made in mps (meters per second).
If a liner was chosen from the LINER MA list, a nominal value for speed of sound in that media will be automatically loaded. If

the actual sound speed rate is known for the pipe liner and that value varies from the automatically loaded value, the value

can be revised.

BSC MENU – BASIC MENU

Page 41