Channel 2 – control output configuration flow only, None, Batch / total – Dynasonics TFX Ultra Transit Time Flow Meters User Manual
Page 63: All alarm outputs are disabled, Multiplier mode, Control 1, Basic tab

Channel 2 – Control Output Configuration Flow Only
Two independent open collector transistor outputs are included with the flow only model. Each output can be configured
independently to alarm for one of the following. See
ALARM OUTPUTS
for output details.
None
Batch / Total
Flow
Signal Strength
Errors
Do w n l o a d
Ca n c el
F i l e Op en . . .
F i l e Sa v e. . .
S y s t e m C o n f i g u r a t i o n
Display
Basic Flow Filtering Output Security
Co n tr o l Ou tp u ts
Channel 2:
50
Off <
Gal/M
350
On>
Gal/M
Mode:
F l o w
Control 1
50
Off <
Gal/M
350
On>
Gal/M
Mode:
F l o w
Mode:
No n e
Control 2
4-20m A / F r eq u en c y
Channel 1:
0
Flow at 4mA / 0Hz:
Gal/M
400
Flow at 20mA / 1KHz:
Gal/M
Calibration/Test
Test
Calibration
4 m A
20 m A
32
3837
Test
4
F l o w
B a tc h / To ta l
F l o w
Si g Str en g th
E r r o r s
Co n tr o l Ou tp u ts
Channel 2:
Figure 45: Channel 2 output choices
OTE:
N
The Batch/Total output is limited to 1Hz maximum pulse rate.
None
All alarm outputs are disabled.
Batch / Total
Multiplier (Value)
This is the value to which the totalizer will accumulate before resetting to zero and repeating the accumulation. This value
includes any exponents that were entered in the BASIC menu as TOTAL E. See
ALARM OUTPUTS
.
50
Multiplier
Mode:
B a tc h / To ta l
Control 1
Figure 46: Control output set for batching
BASIC TAB
Page 63
March 2014