86

DESCRIPTION

The PF_CONTROL command controls the

pass/fail controlling options: “operate”, “output”
and the “stop on output”.

See instrument’s Operator Manual for these

options

The response to the PF_ CONTROL? query
indicates the controlling options of the pass/fail.


COMMAND SYNTAX

PF_ CONTROL

TRACE,,CONTROL,,OUTP
UT,,OUTPUTSTOP,

: = {C1,C2,C3,C4}
: = {START,STOP}
: = {FAIL,PASS}
: = {ON,OFF}

QUERY SYNTAX

PF_ CONTROL?


RESPONSE FORMAT

PF_ CONTROL

TRACE,,CONTROL,,

OUTPUT,,OUTPUTSTOP,


EXAMPLE

The following instruction sets source to channel

1, “operate” to “start”, “output” to “pass” and
“stop on output” to “off”:


Command message:

PFCT TRACE,C1,CONTROL,START,
OUTPUT,PASS,OUTPUTSTOP,OFF

FUNCTION

PF_CONTROL, PFCT

Command /Query