beautypg.com

B&K Precision 2552-2559 - Programming Manual User Manual

Page 13

background image

13

DESCRIPTION

The ATTENUATION command selects the vertical

attenuation factor of the probe. Values of 1, 5, 10, 50,
100, 500, and 1000 may be specified.

The ATTENUATION? Query returns the

attenuation factor of the specified channel.


COMMAND SYNTAX

: ATTeNuation

: = {C1, C2, C3, C4}

: = {1, 5, 10, 50, 100, 500, 1000}

QUERY SYNTAX

: ATTeNuation?


RESPONSE FORMAT

: ATTeNuation


EXAMPLE

The following command sets to 100 the

attenuation factor of Channel 1:

Command message:

C1:ATTN 100

ACQUISITION

ATTENUATION, ATTN

Command /Query