DESCRIPTION

The HOR_MAGNIFY command horizontally

expands the selected expansion trace by a
specified factor. Magnification factors not
within the range of permissible values will
be rounded off to the closest legal value.

If the specified factor is too large for any of
the expanded traces (depending on their
current source), it is reduced to an
acceptable value and only then applied to
the traces. The VAB bit (bit 2) in the STB
register is set when a factor outside the legal
range is specified.

The HOR_MAGNIFY query returns the
current magnification factor for the
specified expansion function.


COMMAND SYNTAX

: Hor_MAGnify

: = {TA, TB, TC, TD}

: = 1 to 50,000,000 The range of
it is related to the current timebase
and the range of the timebase


QUERY SYNTAX

: Hor_MAGnify?


RESPONSE FORMAT

: Hor_MAGnify


EXAMPLE

The following instruction horizontally

magnifies Trace A (TA) by a factor of 5:

Command message:
TA: HMAG 5.00


RELATED COMMANDS

HPOS

DISPLAY

HOR_MAGNIFY, HMAG

Command /Query