DESCRIPTION

The VOLT_DIV command sets the vertical

sensitivity in Volts/div. The VAB bit (bit 2) in
the STB register is set if an out-of-range value is
entered.

The VOLT_DIV query returns the vertical
sensitivity of the specified channel.


COMMAND SYNTAX

: Volt_DIV

:= {C1, C2, C3, C4}

:= 2mV to 10V(or 5V in the CFL
series)

Note: The suffix V is optional.


QUERY SYNTAX

: Volt_DIV?

RESPONSE FORMAT

: Volt_DIV


EXAMPLE

The following command sets the vertical

sensitivity of channel 1 to 50 mV/div:


Command message:
C1: VDIV 50MV

ACQUISITION

VOLT_DIV, VDIV

Command /Query