1 loading samples with a syringe – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 75
3 • Operation and Maintenance
Doc. 065291-01 3/09
63
Setup for an AS-HV Autosampler
For setup information for the AS-HV Autosampler, see the AS-HV Autosampler
Operator’s Manual (Document No. 065125). The manual is provided on the
Dionex Reference Library CD-ROM (P/N 053891).
3.10.1 Loading Samples with a Syringe
1. Make sure the injection port on the ICS-2100 front door (see
) is connected to sample port
S (5)
on the injection valve
(see
Figure 3-10
).
2. Fill the 1 cc syringe (P/N 016388) provided in the ICS-2100 Ship Kit
(P/N 064375) with a calibration standard or sample.
3. Insert the syringe into the injection port on the ICS-2100 front door
4. Verify that the injection valve is in the Load position.
5. Overfill the sample loop with several sample loop volumes. Excess
sample will exit through the injection valve waste line.
6. Leave the syringe in the port.
7. Switch the injection valve to the Inject position (see
Figure 3-10. Injection Valve Connections