B.6.2 status parameter details, B.6.2, Status parameter details – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 233: B • touch screen operation
B • Touch Screen Operation
Doc. 065291-01 3/09
221
2. Touch the down arrow next to the field you want to change. A list of
the available status parameters appears (see
3. Touch the parameter you want to display.
4. When finished, touch the
Apply
button.
B.6.2
Status Parameter Details
Figure B-12. Status Page Parameter List
Status Parameter
Description
Eluent time
The number of hours the eluent supply will last, if the pump runs
continuously at the current flow rate.
Flow rate
The actual flow rate currently output by the pump.
EGC1 conc
EGC2 conc
The actual concentration of eluent currently being generated by the
EluGen cartridge (EGC1 or EGC2).
AUX PS
The actual auxiliary power supply current in mA.
This field replaces the
EGC2 conc
field when the EGC-2 power
supply is configured as an auxiliary power supply.
Suppressor
The actual current applied to the suppressor.
Elapsed time
The number of hours since the last injection.