1 recording the epm serial number in chromeleon – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 182
ICS-2100 Ion Chromatography System
170
Doc. 065291-01 3/09
5.21.1 Recording the EPM Serial Number in Chromeleon
The serial number is printed on a label on the EPM.
1. Open the Server Configuration program (select Start > All
Programs > Chromeleon > Server Configuration).
2. Double-click the ICS-2100 system icon. The properties dialog box for
the ICS-2100 appears.
3. Click the Eluent Generator tab (see
Figure 5-44
).
4. Select the EGC-2 check box.
5. Enter the serial number of the EPM.
6. Verify that the cartridge type is
pH Mod
. If not, make sure the serial
number was entered correctly.
7. Click OK.
Figure 5-64. Eluent Generator Configuration Properties Example