Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 145
5 • Service
Doc. 065291-01 3/09
133
9. Enter the column heater calibration values in Chromeleon or on the ICS-2100
touch screen
DIAGNOSTIC
page as follows.
To enter the calibration values in Chromeleon:
a.
Open the Wellness Panel (see
) and click Calibration
Details.
b. Under Column Heater Calibration Values, enter the Column Heater
Offset and Column Heater Slope values recorded in
.
c.
In the field next to the Download button, verify that Current is selected
and then click Download to download the new calibration values to the
ICS-2100.
d. To ensure the new calibration values are recorded in the ICS-2100
memory, do not turn off the ICS-2100 power for at least 1 minute after
downloading the values.
To enter the calibration values on the ICS-2100 touch screen
DIAGNOSTIC
page:
a.
Go to the
DIAGNOSTIC
page and touch
Column Heater Cal Values
.
b. On the
COLUMN HEATER CAL VALUES
page (see
Figure 5-24
), enter the
Calibrated Offset
and
Calibrated Slope
values recorded in
Figure 5-24. Column Heater Cal Values Page